[Intended Use]
NiTi Brush for peri-implantitis is intended for mechanical debridement and cleaning of titanium dental implant surface contaminated by osseous defect of plague, calculus or any other foreign body materials resulted from peri-implantitis, and NiTi Brush for peri-implantitis is designed for use in dental office by dentists, dental hygienists and trained dental assistants.
[Indication] : Peri-implantitis disease
[Side Effects] : Redness, Discomfort, Swelling
[Contraindications]
- It is contraindicated in young patient with poor attention and it can’t be used in non-implant surgery patients.
- Usage of the device for pregnant, pediatric or elder patient, there is no safety information, please consult to specialist
[Patient Population]
- Ages : adult
- Weight : Multiple
- Race : Multiple
- Patient condition : The NiTi Brush is used for patients who have dental implants.
[Intended Operator]
- Education : Expert trained who can use this medical device.
- Knowledge : Medical Knowledge
a) Read and understand ‘westernized Arabic’ numerals when written in Calibri font
b) Can distinguish: left arm, right arm
c) Understand hygiene
- Language : English or German (Languages offered in the instruction for use)
- Experience : Use by dentists, dental hygienists and trained dental assistants
[Checklist before Use]
- Recommended temperature for operation room is 15℃~30℃.
- The reactions of patient against or incompatibility with titanium or nickel.
- Visual inspection for breakage of package.
- Visually inspect the damage, breakage, deformation, pollution from any foreign material in NiTi Brush
- Sterilization the NiTi Brush according to the appointed sterilization method
- Preparing recommended handpiece for NiTi Brush.
- Preparing sterile saline solution(NaCl) for irrigation and cooling.
- Preparing 3% hydrogen peroxide solution for the multiple use in one patient using one NiTi Brush
- Understanding of instruction for use.
[How to Use]
- Open the outer box and pick the inner plastic box up.
- Open the inner plastic box and pick the NiTi Brush up
- Inspect the NiTi Brush for defects
- Pack the NiTi brush into the sterilization pouch
- Sterilize the NiTi Brush using autoclave.
- Attach the NiTi Brush in sterilized handpiece.
- Position the NiTi Brush to clean the targeted dental implant then clean the dental implant.
- Rinse extensively with sterile saline solution throughout the whole brushing time.
- In case of multiple use in one patient using one NiTi Brush, it is recommended to immerse the NiTi Brush in a 3% hydrogen peroxide solution and rinse with sterile saline before clean or debride another implant.
[How to connect between NiTi brush and Dental Handpiece]
- Recommended handpiece for NiTi Brush is the CE marked rotating implant motor(handpiece).
- Any handpiece that has the deceleration rate of “20:1” is recommended.
- Recommended RPM for operation is 600RPM ~ 1,200RPM.
[Caution]
- Only use by dentist or dental hygienist and trained dental assistants
[Warning]
- NiTi Brush must not be used in patients reporting reactions against or incompatibility with titanium or nickel.
- Do not use the brush in the broken inner plastic box.
- In case of damage, breakage, deformation, pollution from any foreign material in NiTi Brush, do not use it.
- NiTi Brush is non-sterile product.
- NiTi Brush must be sterilized according to the sterilization method & condition of instruction for use.
- After sterilization, use the NiTi Brush without re-packing or re-storage.
- Because of the risk for cross infection from patient to patient, NiTi Brush should be discarded after use.
- NiTi Brush is for single-use only. Do not re-use the NiTi Brush.
- Do not use if expiration date has been exceeded.
- NiTi Brush must be used in accordance with the instructions for use provided by the manufacturer.
[Sterilization Method & Condition]
- Preparing the sterilization of NiTi Brush and sterile the NiTi Brush in accordance with bellow condition.
- Autoclaving Procedure
Sterilization Method |
Autoclave |
Configuration |
Wrap |
Temperature |
134℃ |
Pressure |
2.5bar |
Exposure Time |
15 minutes |
Dry Time |
20 minutes |
- Keep the NiTi Brush in the sterilization pouch (or Sterilization Wrap) to keep it clean until you use it.
SYMBOL |
DEFINITION |
SYMBOL |
DEFINITION |
|
MANUFACTURER |
|
AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY |
|
BATCH CODE (LOT NO.) |
|
CATALOGUE NUMBER (MODEL NAME) |
|
CAUTION |
|
CONSULT INSTRUCTIONS FOR USE |
|
DATE OF MANUFACTURE |
|
DO NOT REUSE |
|
NON STERILE |
|
USE-BY DATE |
|
KEEP AWAY FROM SUNLIGHT |
|
KEEY DRY |
DO NOT USE IF PACKAGE IS DAMAGED |
|
|